Z-0005-2009

Toshiba Infinix fluoroscopy DFP-8000D Digital Radiography System with version 3.40 or higher software...

Canon Medical Systems USA, Inc. / initiated 2008-09-29 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Toshiba Infinix fluoroscopy DFP-8000D Digital Radiography System with version 3.40 or higher software. Controller for Infinix interventional angiography systems on . The reason FDA records is "Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0005-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba Infinix fluoroscopy DFP-8000D Digital Radiography System with version 3.40 or higher software. Controller for Infinix interventional angiography systems.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba issued a letter entitled Important Electronic Product Radiation Warning, September 30, 2008, explaining the issue. The customers were requested to reply to the letter via fax-back form, and a Toshiba service representative would install the corrective action.
Quantity in commerce
129
Distribution
Nationwide
Product codes and lots
Software version 3.40 or higher