Z-1693-2010

Juxta-Lite compression wrap Large Short 28cm Length with Separate Small AFW Catalog 23125017

Circaid Medical Products Inc / initiated 2010-04-20 / Terminated / root cause: manufacturing process

Circaid Medical Products Inc began a recall of Juxta-Lite compression wrap Large Short 28cm Length with Separate Small AFW Catalog 23125017 on . The reason FDA records is "Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1693-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Circaid Medical Products Inc
Product
Juxta-Lite compression wrap Large Short 28cm Length with Separate Small AFW Catalog 23125017
Product code
MDR Binder, Medical, Therapeutic
Reason for recall
Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension level
Root cause tag
manufacturing process
FDA root cause
Process control
Action
CircAid Medical Products Important Device Product Recall letter dated Apr 20, 2010 was mailed to consignees describing problem, listing affected product and distribution dates. Customers were instructed to locate and return these Juxta-Lite Systems from stock and notify their customers A Retum Authorization number and contact name was provided. this listing for product returns.
Quantity in commerce
372 total, all sizes
Distribution
Worldwide distribution: USA, Japan, Canada, Netherlands
Product codes and lots
All units sold between 2/1/10 and 4/14/10.