Circaid Medical Products Inc appears in FDA device records in San Diego, CA. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 19 recalls for this company.

FDA records held for this company

510(k) clearances
0
PMA records
0
Recall records
19
Registered establishments
0

510(k) clearance history

No 510(k) clearance records for this company in the ingested FDA data.

No 510(k) clearance records in the ingested FDA data.

Registered establishments

No FDA establishment registrations recorded against this company.

FDA product codes

No FDA product codes recorded against this company.

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-1702-2010Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Separate Medium AFW Catalog 23236117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1701-2010Juxta-Lite compression wrap XL Long 33cm Length with Separate Medium AFW Catalog 23236017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1700-2010Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Separate Small AFW Catalog 23226117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1699-2010Juxta-Lite compression wrap XXL Short 28cm Length with Separate Medium AFW Catalog 23137017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1698-2010Juxta-Lite compression wrap XL Full Calf Short 28cm Length with Separate Medium AFW Catalog 23136117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1697-2010Juxta-Lite compression wrap Large Short 28cm Length with Separate Medium AFW Catalog 23135017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1696-2010Juxta-Lite compression wrap Medium Short 28cm Length with Separate Medium AFW Catalog 23134017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1695-2010Juxta-Lite compression wrap XL Full Calf Short 28cm Length with Separate Small AFW Catalog 23126117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1694-2010Juxta-Lite compression wrap XLShort 28cm Length with Separate Small AFW Catalog 23126017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1693-2010Juxta-Lite compression wrap Large Short 28cm Length with Separate Small AFW Catalog 23125017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1692-2010Juxta-Lite compression wrap Medium Short 28cm Length with Separate Small AFW Catalog 23124017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1691-2010Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Anklet Catalog 23036117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1690-2010Juxta-Lite compression wrap XL Long 33cm Length with Anklet Catalog 23036017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1689-2010Juxta-Lite compression wrap XXLShort 28cm Length with Anklet Catalog 23027017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1688-2010Juxta-Lite compression wrap Large Full Calf Short 28cm Length with Anklet Catalog 23025117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1687-2010Juxta-Lite compression wrap XL Full Calf Short 28cm Length with Anklet Catalog 23026117Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1686-2010Juxta-Lite compression wrap XL Short 28cm Length with Anklet Catalog 23026017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1685-2010Juxta-Lite compression wrap Medium Short 28cm Length with Anklet Catalog 23024017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated
Z-1684-2010Juxta-Lite compression wrap Large Short 28cm Length with Anklet Catalog 23025017Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension levelTerminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.