Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Separate Medium AFW Catalog 23236117
Circaid Medical Products Inc began a recall of Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Separate Medium AFW Catalog 23236117 on . The reason FDA records is "Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1702-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Circaid Medical Products Inc
- Product
- Juxta-Lite compression wrap XL Full Calf Long 33cm Length with Separate Medium AFW Catalog 23236117
- Product code
- MDR Binder, Medical, Therapeutic
- Reason for recall
- Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension level
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- CircAid Medical Products Important Device Product Recall letter dated Apr 20, 2010 was mailed to consignees describing problem, listing affected product and distribution dates. Customers were instructed to locate and return these Juxta-Lite Systems from stock and notify their customers A Retum Authorization number and contact name was provided. this listing for product returns.
- Quantity in commerce
- 372 total, all sizes
- Distribution
- Worldwide distribution: USA, Japan, Canada, Netherlands
- Product codes and lots
- All units sold between 2/1/10 and 4/14/10.