Voiceware clip software update to certain Horizon Medical Imaging Systems. McKesson Medical Imaging Company...
Mckesson Medical Immaging began a recall of Voiceware clip software update to certain Horizon Medical Imaging Systems. McKesson Medical Imaging Company. Richmond, Canada V6X 3G5. Intended to receive... on . The reason FDA records is "Voice Clip issue: When the user is recording voice clips in Horizon Rad Station, there is a potential for voice clips to be saved to the incorrect study of the same patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1735-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Change Healthcare Canada Company
- Product
- Voiceware clip software update to certain Horizon Medical Imaging Systems. McKesson Medical Imaging Company. Richmond, Canada V6X 3G5. Intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video and associated medical information from various medical imaging systems.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Voice Clip issue: When the user is recording voice clips in Horizon Rad Station, there is a potential for voice clips to be saved to the incorrect study of the same patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- McKesson Medical Imaging Company issued an "Advisory Notice" dated May 7, 2010. Consignees were notified of the affected product and advised to contact McKesson Customer Support department to review their sites current configuration and install the appropriate product software update(s). For further information, contact McKesson Medical Imaging Customer Support at 1-800-663-2533. On June 19, 2010 the recall was amended to include the following actions to follow: Instead of highlighting the anchor study in the Patient Portfolio Sidebar before saving t he voice clip in the Patient Portfolio Sidebar, McKesson recommends that you save the voice clip to the desired study first, before browsing through other documents. A software update that prevents the problem from occurring is currently under development for all affected releases and will be available in approximately three months. Consignees will be notified when the software update is ready to be installed on their system.
- Quantity in commerce
- 100 units
- Distribution
- Distributed Nationwide (AK, AL, AR, AZ, CA, CO, FL, GA, IN, IA, IL, KY, LA, MA, MD, MI, NC, ND, NE, NM, NY, NV, OH, OR, PA, RI, SC, TN, TX, WY, VA and WV) and Canada
- Product codes and lots
- Horizon Medical Imaging (HMI) versions: 4.5, 4.6.1, 5.0 without PTF-RR310-601, 11.0.3, 11.0.4, 11.0.5, 11.0.6, 11.0.6 FR EXP2 without PTF-HMI11-649, 11.0.7, 11.0.8 without PTF-HMI11-615, 11.5 and 11.5.1 without PTF-HMI115-613. On June 29, 2010 the recall was amended to include: All installations of Horizon Medical imaging (HMI) 11.0.6, 11.0.8 and 11.5.1