Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating...
Toshiba American Medical Systems Inc began a recall of Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with... on . The reason FDA records is "Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2981-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies.
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage shall not exceed 60V after 1 second from taking off the AC plug.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Toshiba America Medical Systems (TAMS) sent an "URGENT:MEDICAL DEVICE CORRECTION' letter dated June 2, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to be careful not to touch the contacts or prongs on the AC plug within two seconds after unplugging the system from a wall socket. This precaution is to be adhered to until a Toshiba Service Representative replaces the power supply. Customers are to complete and return a Customer Reply Form and fax to 877-349-3054. Customers with questions regarding the notification should contact the Director of Regulatory Affairs at (800) 421-1968 or call their local Toshiba Representative at (800) 521-1968.
- Quantity in commerce
- 78
- Distribution
- Nationwide Distribution
- Product codes and lots
- 4899 5131, 5132, 5133, 5134, 5224, 5139, 5260, 5261, 5369, 5256, 5367, 4883, 4884, 5370, 5374, 5375, 5264, 5257, 5135, 5223, 5379, 5262, 6283, 4886, 5258, 5382, 5383, 5384, 5385, 5216, 4887, 4897, 4898, 5027, 5380, 4982, 5026, 5137, 5138, 5219, 5220, 4979, 5465, 4900, 5466, 4978, 5377, 5136, 5376, 5072, 5071, 5221, 5215, 4981, 5371, 5372, 5373, 5029, 5030, 5068, 5069, 5070, 5222, 4980, 4885, 5259, 5028, 4896, 5368, 5217, 5218, 5140, 5265, 5364, 5365, 5366, 5381, 5467, 5468, 5469, 5470, 5471, 5472, 5473, 5474