Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner
Toshiba American Medical Systems Inc began a recall of Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner on . The reason FDA records is "Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1257-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Toshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
- Quantity in commerce
- 44 units distributed between 09/2007 - 07/2010
- Distribution
- Nationwide distribution.
- Product codes and lots
- Affected Serial No.: 2DA0872038, 2DA0962116, 2DA0862034, 2DA09Z2161, 2DA07X2008, 2DA1012165, 2DA08X2072, 2DA09Y2151, 2DA0842023, 2DA07X2005, 2DA1012168, 2DA08X2067, 2DA0882053, 2DA0862036, 2DA08Z2092, 2DA0882051, 2DA09Z2162, 2DA1072211, 2DA09X2143, 2DA08Y2077,2DA09X2139, 2DA08X2070, 2DA0982133 2133, 2DA0842026, 2DA0882052, 2DA0972123, 2DA08Y2080, 2DA0882048, 2DA08X2063, 2DA09Y2148, 2DA1012170, 2DA08X2071, 2DA1012169, 2DA0952115, 2DA0982131, 2DA07Y2010, 2DA08X2061, 2DA1012167, 2DA0882049, 1DA0962006, 1DA09Y2014, 1DA0992011, 1DA0972007, 1DA0972008, 2DA1012166, 2DA09Z2158, 2DA1072212, 2DA1012174, 2DA0792004, 2DA1062208, 2DA0792002, 1DA1062015, 1DB1072012