Aplio Artida Ultrasound Diagnostic System SSH-880CV. Toshiba American Medical Systems, Inc. A diagnostic...
Toshiba American Medical Systems Inc began a recall of Aplio Artida Ultrasound Diagnostic System SSH-880CV. Toshiba American Medical Systems, Inc. A diagnostic ultrasound system on . The reason FDA records is "Toshiba America Medical Systems, Inc. is initiating a field correction on the Aplio Artida Ultrasound Diagnostic System because the panel operation for the Aplio Artida may become disabled (i.e., system may hang up)...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2897-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Aplio Artida Ultrasound Diagnostic System SSH-880CV. Toshiba American Medical Systems, Inc. A diagnostic ultrasound system.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Toshiba America Medical Systems, Inc. is initiating a field correction on the Aplio Artida Ultrasound Diagnostic System because the panel operation for the Aplio Artida may become disabled (i.e., system may hang up) when processing to display patient information (heart rate, time, VCP counter, ect.).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Toshiba America Medical Systems, Inc. sent out an "Urgent Medical Device Correction" dated December 21, 2010 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were notified that Toshiba will issue a field modification to correct the software, and if the problem occurs, they were instructed to shutdown the system holding down the power supply button for 5 seconds for 5 seconds and then restart the system. A form to send back to Toshiba is included. For questions on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
- Quantity in commerce
- 46 units
- Distribution
- Worldwide Distribution: (USA) Nationwide distribution including the states of MD, OH, PA, NY, MA, GA, AL, TX, IL, WI, IA, MN, KY, MT, OR, LA, and CA; and the country of Puerto Rico.
- Product codes and lots
- Last 4 of serial number: 2342, 2480, 2073, 2120, 2219, 2270, 2271, 2268, 2389, 2390, 2297, 2206, 2033, 2107, 2087, 2311, 2281, 2273, 2340, 2341, 2313, 2272, 2070, 2335, 2347, 2348, 2360, 2361, 2048, 2280, 2031, 2088, 2076, 2077, 2105, 2334, 2220, 2260, 2205, 2072, 2074, 2079, 2218, 2103, 2262, 2069, 2106, 2119, 2124, 2204, 2405, 2104, 2032, 2075, 2080, 2108, 2261, 2312, 2078, 2071, 2123, 2296, 2282, 2049, and 2056.