Z-2896-2011

Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A...

Canon Medical Systems USA, Inc. / initiated 2011-01-07 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning on . The reason FDA records is "Toshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2896-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Toshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to failure of the optical conversion connector, resulting in a break of communication between units.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba America Medical Systems, Inc. sent an "Urgent Medical Device Correction" letter dated January 4, 2011 to all affected customers. The letter provides the customers with an explanation of the problem and the firms plan to install preventive measures and provided them with a Customer Reply form to send Toshiba. A representative will contact the customers for an appointment to install the measures to their CT system. For question on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
Quantity in commerce
5
Distribution
(USA) Nationwide Distribution including the states of WI, MO, TN, IA, AL, AZ, TX, and GA.
Product codes and lots
1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019