Z-2399-2011

ES0101-45, AEM Disposable Curved Scissors, 1/2", 45cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting...

Encision, Inc. / initiated 2011-04-07 / Terminated / root cause: packaging

Encision, Inc. began a recall of ES0101-45, AEM Disposable Curved Scissors, 1/2", 45cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures on . The reason FDA records is "Sterility may be compromised due to holes in packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2399-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encision, Inc.
Product
ES0101-45, AEM Disposable Curved Scissors, 1/2", 45cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Sterility may be compromised due to holes in packaging.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Encision, Inc. sent an Important Safety Notice letter dated April 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to either carefully inspect any remaining pouches for readily visible holes in the Tyvek pouch, or return the affected lots to Encision for inspection. Customers were also asked to fill in and fax back the attached Confirmation to 303-444-2693.
Quantity in commerce
44 Ten-Count Boxes
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia and New Zealand
Product codes and lots
Lot # QDI, QJJ, RAF